Error 326 bv pulsera
December 04, 2018 09:56
This problem 326/327 only occurs at startup. Never during normal operation. Incorrect init. of a ADC circuit on XGC pulse bd. once again replace SHA4 gen cont. per FCO 718 00 018...Click here to read more
Error 326 bv pulsera
December 04, 2018 09:56
self test of the RA Generator control board. Replace SHA4. Now I have seen this before and if you unplug the umbilical cord and machine. hook back up and reboot it might clear. There is a Philips service note on that error. cant...Click here to read more
Artefacts on philips Optimus 65
November 12, 2018 10:58
I agree, please ALWAYS post the basics in the first part of a post. 1. real system type and software version. You did not say say what the artifacts looked like. Before you do gain and pixel you should do tube adaption first. Do...Click here to read more
GE paint codes
October 20, 2018 01:12
almost forgot...I used a boat repair/restoration shop for a CT front cover. Looked like new and they only charge 450.00...Click here to read more
GE paint codes
October 20, 2018 01:12
They have to be able to withstand bleach and peroxide. and other cleaning chamicals. There are some auto paints that may work but in a nutshell paint with single stage polyurethane paint, but some repair work is being done in base...Click here to read more
GE Mammocom Software
October 19, 2018 09:45
E01 and 05 are user codes...what is the 6 dig. code from the system error log" Your IDC is not communicating properly to generator. check your cabling. check the display on the rear of your IDC computer. It should be in the syst...Click here to read more
TCOP
October 06, 2018 12:50
To get this from Philips you will 1. need official Philips training on that product. 2. Sign at a minimum a "tech support" agreement. 3. with a Coop agreement it is "no charge" not sure on the support. Once you get IST security...Click here to read more
RFAMP 53-S30-64 --- FAULT 12
September 13, 2018 01:28
I missed the S30 part. Yep that is the single unit. heavy and hard to get to the power connections on the back depending on site layout. There was an FCO that resolved some issues but I cant remember what it solved..ill have to d...Click here to read more
RFAMP 53-S30-64 --- FAULT 12
September 13, 2018 01:27
ahhh..that makes sense... S30 is the newer Achiva style. That is the heavy one piece unit. sits in the bottom and is one heavy pig. You can check with All MKS amplifiers are built by IPI Group LLC in Colorado Springs, CO. 3337 N ...Click here to read more
Logiq P5 external VGA monitor
September 11, 2018 09:27
as a user you don't have the "privileges" in that user account to get into the OS. maybe youll get lucky...plug it in and maybe it is set already for slave monitor....Click here to read more
RFAMP 53-S30-64 --- FAULT 12
September 11, 2018 09:27
arcing tubes will cause that error. however...if that has been going on for a while then the power supply could be damaged. Or it is the power supply. you can replace the lower PS deck without the upper. There is a separate P/N f...Click here to read more
Philips Intera 1.5T
April 28, 2018 11:45
depends on your software/hardware release level. Need more than just Intera or its all guess work....Click here to read more
Looking for Rel.Bulletin R11.1.4 for MR Philips SW Install
April 23, 2018 04:08
If you get the rel. bulletin, don't forget that you have about 6 service packs and multiple security fixes to load after that....Click here to read more
Philips MRI Gradien power supply - Intera 1.5T
April 18, 2018 09:58
what version of system software are you running? ...Click here to read more
Philips MRI Gradien power supply - Intera 1.5T
April 18, 2018 09:58
did you check the Cap bank?...Click here to read more