Over 1650 Total Lots Up For Auction at Two Locations - PA 07/14, NJ Cleansweep 07/16

stats
DOTmed text ad

Probo Medical | We Buy, Repair, and Sell New and Used Ultrasound Equipment

We provide sales, service, rental, and repair for ultrasound systems, probes, and C-arms. We have the largest inventory in the world. Contact us about our special offers on certified refurbished equipment! Call: 317-759-9210 Email: sales@probomedical.com.

stats
stats
DOTmed text ad

North America's Ultimate Probe Repair Lab & Certified Refurbished OEM Probes in Stock

North America's trusted source for ultrasound probe repair and certified refurbished OEM probes. Free evaluations, 1-year warranty, fast turnaround for GE, Philips, Sonosite, Canon, Toshiba, and more. Call (224) 265-7640 / Sales@Lenard.Healthcare

DOTmed Certified stats
Mako meter by RTI Group
Mako meter by RTI Group
New Product Showcase
spacer
Por página

Listas

Service For GE HEALTHCARE Repair Voluson E10 (RFM 323). Repair Voluson E10 (RFM 323).

Faulty Phenomenon Malfunction: during operation of the device in Doppler mode in the area of interest, colored noise appears. Troubleshooting and Solutions Malfunction: during operation of the... Veja mais

June 30  

Service For GE HEALTHCARE Repairing Logiq e. Repairing Logiq e.

Faulty Phenomenon Malfunction: The cursor on the screen system did not respond to movements of the trackball horizontally (left and right), which complicated the control of the device. ... Veja mais

June 30  

Service For PHILIPS MEDICAL Repairing the Philips Epiq 5 Repairing the Philips Epiq 5

Faulty Phenomenon Malfunction: when turning on the device during the operation process, an error with the diagnostic code DiagCode:200 appears. J49P. 3FXS. 5. No possibility to work, requires... Veja mais

June 30  

Service For GE HEALTHCARE Repair Voluson E10 BT16. Repair Voluson E10 BT16.

Faulty Phenomenon Malfunction: the issue has a progressive nature, that is, earlier during the scanning process, an error occurred sporadically with the frequency of "Error: Aurora communication... Veja mais

May 25  

Service For GE HEALTHCARE Repairing GE Voluson E10 equipment. Repairing GE Voluson E10 equipment.

Faulty Phenomenon Malfunction: During the examination, artifacts appear in the form of bright spots, morgania, etc. With each examination, the artifacts increase. Solution: The malfunction was... Veja mais

May 25  

Service For GE HEALTHCARE Repair equipment GE Logiq E9 R6. Repair equipment GE Logiq E9 R6.

Faulty Phenomenon Malfunction: when turning on, an error appears, associated with overheating of the device solution: the TO of the device was performed, and the MRX192 board was also repaired. ... Veja mais

June 30  

Service For PHILIPS Repair medical equipment Epiq 5, Epiq ,7 Repair medical equipment Philips Epiq 5, Epiq 7

Faulty Phenomenon Malfunctions: artifacts in B, Color modes, error 005. Board Acquisition Module faulty in all devices solution: repair the Acquisition Module board within 3 working days. ... Veja mais

June 30  

Service For PHILIPS MEDICAL Repairing ultrasound equipment Philips HD11 and HD15. Philips HD11 and HD15.

Faulty Phenomenon Malfunction: The device does not respond to the power button, the indicators on the Power Supply block do not light up. Faulty radio components in PFC (input section) have been... Veja mais

June 30  

Service For GE HEALTHCARE Repair ultrasound GE Logiq S8 R2. Repair ultrasound GE Logiq S8 R2.

Faulty Phenomenon Malfunction: Based on the results of the service testing, the board BF192 (5357234-20) is faulty. The malfunctioning channel has been identified as the cause of the device's... Veja mais

May 25  

Service For PHILIPS Repairing ultrasound iU22 (E. 3, software 6.3.6). Philips iU22

Faulty Phenomenon Malfunction: the device responds to the power button but fails to start the ultrasound application. A malfunction was detected in the hard disk. The hard disk was replaced and the... Veja mais

June 30  

Service For PHILIPS Repairing the Epiq Elite involves addressing various technical iss Repairing the Philips Epiq Elite

Faulty Phenomenon Problem description: when turning on systems, an error 005 (005. 3DR4. 692A. 16) appears. Troubleshooting and Solutions Description of the problem: when turning on the system, an... Veja mais

June 30  

Service For GE HEALTHCARE Repair GE Voluson E8 (BT12). Repair GE Voluson E8 (BT12).

Faulty Phenomenon A stripe appeared on the display of the RST board (PN: KTZ196358, KTZ280122). RST - impulse transmission board, 16 pieces are installed in the device (1 faulty). Sent the board... Veja mais

June 30  

Service For GE HEALTHCARE Repair GE Logiq E10 (version R3). Repair GE Logiq E10 (version R3).

Faulty Phenomenon GE Logiq E10: A malfunction was detected in the TX board, an element that controls the switching on/off of high voltage for supply to the acoustic module has failed. The device is... Veja mais

June 30  

Service For GE HEALTHCARE Repair Beamformer of Voluson E10 BT16. Repair Beamformer of Voluson E10 BT16.

Faulty Phenomenon Beamformer consists of RFM201, RFM221 (PN: KTZ303915, KTZ303916); RSX10, RSX20 (KTZ303250, KTZ303054). Error: Aurora communication failed; when turning on the device, the image... Veja mais

June 30  

Service For GE HEALTHCARE Repair GE Logiq E9 R6. Repair GE Logiq E9 R6.

Faulty Phenomenon Repair block of power supply Main Power Supply (PN: 5205054). Malfunction: absence of 48V for the computer power supply block and therefore the device does not turn on. ... Veja mais

June 30