Faulty Phenomenon Malfunction: during operation of the device in Doppler mode in the area of interest, colored noise appears. Troubleshooting and Solutions Malfunction: during operation of the... Veja mais
Faulty Phenomenon Malfunction: The cursor on the screen system did not respond to movements of the trackball horizontally (left and right), which complicated the control of the device. ... Veja mais
Faulty Phenomenon Malfunction: when turning on the device during the operation process, an error with the diagnostic code DiagCode:200 appears. J49P. 3FXS. 5. No possibility to work, requires... Veja mais
Faulty Phenomenon Malfunction: the issue has a progressive nature, that is, earlier during the scanning process, an error occurred sporadically with the frequency of "Error: Aurora communication... Veja mais
Faulty Phenomenon Malfunction: During the examination, artifacts appear in the form of bright spots, morgania, etc. With each examination, the artifacts increase. Solution: The malfunction was... Veja mais
Faulty Phenomenon Malfunction: when turning on, an error appears, associated with overheating of the device solution: the TO of the device was performed, and the MRX192 board was also repaired. ... Veja mais
Faulty Phenomenon Malfunctions: artifacts in B, Color modes, error 005. Board Acquisition Module faulty in all devices solution: repair the Acquisition Module board within 3 working days. ... Veja mais
Faulty Phenomenon Malfunction: The device does not respond to the power button, the indicators on the Power Supply block do not light up. Faulty radio components in PFC (input section) have been... Veja mais
Faulty Phenomenon Malfunction: Based on the results of the service testing, the board BF192 (5357234-20) is faulty. The malfunctioning channel has been identified as the cause of the device's... Veja mais
Faulty Phenomenon Malfunction: the device responds to the power button but fails to start the ultrasound application. A malfunction was detected in the hard disk. The hard disk was replaced and the... Veja mais
Faulty Phenomenon Problem description: when turning on systems, an error 005 (005. 3DR4. 692A. 16) appears. Troubleshooting and Solutions Description of the problem: when turning on the system, an... Veja mais
Faulty Phenomenon A stripe appeared on the display of the RST board (PN: KTZ196358, KTZ280122). RST - impulse transmission board, 16 pieces are installed in the device (1 faulty). Sent the board... Veja mais
Faulty Phenomenon GE Logiq E10: A malfunction was detected in the TX board, an element that controls the switching on/off of high voltage for supply to the acoustic module has failed. The device is... Veja mais
Faulty Phenomenon Beamformer consists of RFM201, RFM221 (PN: KTZ303915, KTZ303916); RSX10, RSX20 (KTZ303250, KTZ303054). Error: Aurora communication failed; when turning on the device, the image... Veja mais
Faulty Phenomenon Repair block of power supply Main Power Supply (PN: 5205054). Malfunction: absence of 48V for the computer power supply block and therefore the device does not turn on. ... Veja mais