Over 1650 Total Lots Up For Auction at Two Locations - PA 07/14, NJ Cleansweep 07/16

stats
DOTmed text ad

North America's Ultimate Probe Repair Lab & Certified Refurbished OEM Probes in Stock

North America's trusted source for ultrasound probe repair and certified refurbished OEM probes. Free evaluations, 1-year warranty, fast turnaround for GE, Philips, Sonosite, Canon, Toshiba, and more. Call (224) 265-7640 / Sales@Lenard.Healthcare

DOTmed Certified stats
stats
DOTmed text ad

Probo Medical | We Buy, Repair, and Sell New and Used Ultrasound Equipment

We provide sales, service, rental, and repair for ultrasound systems, probes, and C-arms. We have the largest inventory in the world. Contact us about our special offers on certified refurbished equipment! Call: 317-759-9210 Email: sales@probomedical.com.

stats
Mako meter by RTI Group
Mako meter by RTI Group
New Product Showcase
spacer
Por página

Listas

Service For SIEMENS Array Shim Device for Magnetom ESSENZA Shim service for Magnetom ESSENZA MRI Tools

🧲 M.R. Tech – Complete Siemens MRI Solutions We provide one of the most comprehensive ranges of Siemens MRI solutions in the market, offering genuine and compatible parts for MRI scanners, coils,... Veja mais

June 17  
Francisco Neto / M.R. Tech
telefone: +55 5511998770846

Service For PHILIPS MEDICAL Repairing the ultrasound device Philips Epiq 5. Repairing the ultrasound device Philips Epiq 5.

Faulty Phenomenon Malfunction: On all connected sensors, a part of the display was missing, visualization was incomplete. Troubleshooting and Solutions Malfunction: On all connected sensors, a... Veja mais

June 30  

Service For GE HEALTHCARE Repairing the ultrasound device GE ML6-15-D. Repairing the ultrasound device GE ML6-15-D.

Faulty Phenomenon Replacement of the acoustic lens (photo before replacement, inside the post - after... Veja mais

June 30  

Service For GE HEALTHCARE Repairing the ultrasound device GE RM6C. Repairing the ultrasound device GE RM6C.

Faulty Phenomenon Restoration of the sealing of the piston (problem: leakage of oil was detected under the piston). Troubleshooting and Solutions Restoration of the bellows seal (problem: an oil... Veja mais

June 30  

Service For PHILIPS Repairing the Epiq 5G device. Repairing the Philips Epiq 5G device.

Faulty Phenomenon Malfunction: during operation with any detector, artifacts and artifacts appear in the form of white stripes solution: repair plate Acquisition Module. Troubleshooting and... Veja mais

June 30  

Service For GE HEALTHCARE Updating the software version and iron on the Voluson E8 device. the Voluson E8 device

Faulty Phenomenon The procedure for updating the software version and hardware on Voluson E8 (versions BT10 to BT13) has been carried out. To install the software version BT13, a replacement BackEnd... Veja mais

June 30  

Service For GE HEALTHCARE Updating the software version and iron on the Voluson E8 device. the Voluson E8 device

Faulty Phenomenon The procedure for updating the software version and hardware on Voluson E8 (versions BT10 to BT13) has been carried out. To install the software version BT13, a replacement BackEnd... Veja mais

June 30  

Service For GE HEALTHCARE Repair device GE Philips Epiq 5. Repair device GE Philips Epiq 5.

Faulty Phenomenon Malfunction: error 607 indicates a malfunction in the data base solution: after restarting the PO device, it worked. Troubleshooting and Solutions Malfunction: error 607... Veja mais

June 30  

Service For GE HEALTHCARE GE Voluson E8 BT21 device repair. GE Voluson E8 BT21 device repair.

Faulty Phenomenon Malfunction: During the examination, echo-imaging disappears and everything turns into white color. Solution: repair RFM323 plate. Troubleshooting and Solutions Malfunction:... Veja mais

June 21  

Service For GE HEALTHCARE Repair device Voluson E8 BT13.5. Repair device Voluson E8 BT13.5.

Faulty Phenomenon Malfunction: when connecting any sensor to any connector of the sensor (error PSS) solution: repair of the sensor connection plate RTF was carried out. Troubleshooting and... Veja mais

June 30  

Service For GE HEALTHCARE Repair device GE Voluson E10 B21. Repair device GE Voluson E10 B21.

Faulty Phenomenon Malfunction: during operation, a blue screen appears on the device. Sometimes the device does not turn on and the power button flashes yellow. Solution: repair the power supply... Veja mais

June 30  

Service For GE HEALTHCARE GE Voluson E8 BT13.5 device overhaul. GE Voluson E8 BT13.5 device overhaul.

Faulty Phenomenon Malfunction: during operation of the device, a break occurred and an error Error on Power management (error in the power supply block) appeared. Solution: repair the RFM and RSX... Veja mais

June 30  

Service For PHILIPS MEDICAL Service Logiq P9. 1. Check the device for visible damage and ensure th Service Logiq P9. 1.

Faulty Phenomenon Repair flat MDC Assy (PN 5552220-2, 5552220-3) of GE Logiq P9 malfunction: when turning on the device it displayed the error "System error", and also when connecting the probe it... Veja mais