OCULUS Pentacam HR – High-Resolution Scheimpflug Anterior Segment Tomography
Core Technology
Uses a rotating Scheimpflug camera (360°) paired with a blue LED light source (475 nm, UV‑free) to perform Cornea Fine Scan mode, capturing 100 high-resolution images in only 2 seconds—ideal for fast, non-contact anterior segment imaging.
Feature
Specification
Camera Type
Digital CCD Scheimpflug camera
Scan Speed
100 images / 2 secondsÂ
Curvature Measurement Range
3–38 mm (~9–99 Diopters)
Precision & Reproducibility
± 0.1 D for both curvature and pachymetryÂ
Operating Distance
80 mm (3.1 in)Â
Physical Dimensions (W × D × H)
275 × 320–400 × 500–530 mm (10.8 × 12.6–15.7 × 19.7–20.9 in)Â
Weight (measuring head)
10.6 kg (~23.4 lbs)Â
Power Consumption
Max 42 WÂ
Recommended PC Specs
Intel® Coreâ„¢ i5, 1 TB HDD/SSD, 8 GB RAM, Windows® 10 Pro or newer, USB interface, VESA mountÂ
Advanced Clinical Features
Full Anterior & Posterior Corneal Tomography: Maps elevation, curvature, power distribution, pachymetry, and anterior chamber angle/depth/volume
Early Keratoconus Detection via Belin/Ambrosio Enhanced Ectasia Display, accompanied by ABCD staging progression charts
Corneal Wavefront & Optical Densitometry: Includes Zernike aberration decomposition, densitometry for cataract opacity staging (PNS)
Contact Lens Fitting Assistant with fluorescein simulation and virtual-eye module
DICOM/PACS Integration, 3D anterior chamber analysis, and dual software license included
Typical Clinical Applications
Pre-operative planning for LASIK, cataract, and IOL implantation
Keratoconus screening and progression monitoring
Evaluation of corneal thickness distribution, asymmetry, and epithelial profile
Diagnosis of anterior segment pathologies, covering iris, lens opacities, and angle measurements